Model 98-1 and Model 98-2 (one and two test positions, respectively) meet the requirements ASTM D1881 Foaming Tendency of Engine Coolants test method.
The 12 inch diameter borosilicate glass jar is temperature controlled at 88º C with a stability and uniformity of ± 0.5º C. The temperature controller has an over temperature protection in the event of primary controller failure. A float also protects by shutting down the heater if the water level drops.
A ball in tube flow meter with a precision control valve is provided for each test position. The air flow rate is controlled at 1000 ± 25 mL/min.
98-1
|
Heated liquid test bath with 1 flowmeter, 1 test position |
98-2
|
Heated liquid test bath with 2 flowmeters, 2 test positions |
|
EC-1
|
Test Assembly- 1 each: cylinder, diffuser, air delivery tube, lead washer, stopper |
EC-2
|
Certified spherical alumina gas diffuser |
EC-3
|
Graduated 500 mL cylinder with collar |
FM-5
|
Four position test cylinder storage rack |
FM-14/15
|
Air supply pump assembly with Air drying tower (FM-14+ FM-15) (115 Volts) |
FM-14/15A
|
Air supply pump assembly with Air drying tower (FM-14+ FM-15A) (220 Volts) |
|
|
Diffuser Certification Apparatus |
FM-18D
|
Digital indicating flow rate, certified “U“ tube manometer |
FM-18PC
|
PC controlled apparatus (visual detection) |
FM-19PC
|
PC controlled apparatus (automatic detection) |
|