Test Method: D2386
Automated Freezing Point Analyzer
• Fully Conforms to the Manual ASTM D2386 method
• No Cryostat required
• No Hazardous alcohol required
• Compact design
• Fully Automated
• Self Contained
Click for Manual Freeze Point Apparatus
Click for ALL AUTOMATED models



The Automated Freezing Point Analyzer fully conforms to method ASTM D2386. All dimensions are specified by D2386. The sample is agitated by a wound wire which is driven by a magnetically operated motor. The sample is poured into a receiver. The sample is automatically fed into the test tube, the start button is pressed, and the test begins. The bath is then cooled at a rate such that the sample cooling rate equals the cooling rate of the manual method.

When the test is completed, the apparatus is automatically warmed, the old sample is gravity drained from the test tube (by actuation of a solenoid valve), the fresh sample flushes the test tube and the new test is started.

Fiber optics reflect light from a C shaped mirror immersed in the test sample. When reduced light is detected, the first freezing point is determined. The sample is warmed and the fiber optics detect the melting point.

Automated Freezing Point Schematic

Available Models
Integrated Automated Analyzer with Direct Refrigeration (Dry Bath) Unit.
Part No.
Description
DR4-10
1 Cloud Point automated test position
DR4-11
2 Cloud Point automated test positions
DR4-12
1 Cloud Point + 1 Pour Point automated test positions
DR4-14
1 Cloud Point + 1 CFPP automated test positions
DR4-15
1 Cloud Point + 1 Freeze Point automated test positions
DR4-20
1 Pour Point automated test position
DR4-22
2 Pour Point automated test positions
DR4-24
1 Pour Point + 1 CFPP automated test positions
DR4-25
1 Pour Point + 1 Freeze Point automated test positions
DR4-40
1 CFPP automated test position
DR4-44
2 CFPP automated test positions
DR4-45
1 CFPP + 1 Freeze Point automated test positions
DR4-50
1 Freeze Point automated test position
DR4-55
2 Freeze Point automated test positions

Accessories
Part No.
Description
DR-CP
Test Jar for cloud, mirrored bottom
DR-PP
Test Jar for pour point
DR-CF
Test Jar for CFPP
DR-FP
Test tube for Freeze point
CP-2
O-Ring (one per test jar), dozen (Freeze Point test tube does not require o-ring)
CF-11
Brass filter assembly (as per ASTM D6371 Fig. 7 & 8). Includes CF-22
CF-11A
Filter disk/screen for CFPP (as per Section 6.1.8.1.c of D6371), dozen
CF-11B
Replacement “O” rings for filter screen (large), dozen for CFPP
CF-22
Brass Filter Holder with CF-11A filter screen for CFPP
DR-AP
Aspiration pipette for CFPP
CF-12A
Replacement “O” rings for pipette (small), dozen for CFPP
CF-17
Vacuum pump (as per Section 6.1.12) (115 volt) for CFPP
CF-17A
Vacuum pump (as per Section 6.1.12) (220 volt) for CFPP
Note: Vacuum pump required if suitable vacuum source unavailable. Supports up to 6 positions
DR4-TP
Spare sample temperature probe. Suitable for CFPP, pour, and/or freeze point
DR4-TP-Cloud
Spare sample temperature probe. Suitable for cloud point (spring loaded)
DR-Kit-CP
Recommended spare accessories for automatic Cloud, per position
Includes: (6) DR-CP, (1) CP-2
DR-Kit-PP
Recommended spare accessories for automatic Pour, per position
Includes: (6) DR-PP, (1) CP-2
DR-Kit-CF
Recommended spare accessories for automatic CFPP, per position
Includes: (6) DR-CF, (1) CP-2, (2) CF-11A, (1) CF-11B, (2) DR-AP, (1) CF-12A
Note: Vacuum pump, if needed, ordered separately
DR-Kit-FP
Recommended spare accessories for automatic Freeze, per position
Includes: (6) DR-FP


Specifications are subject to change without notice.
Specifications subject to change without notice. Lawler Manufacturing Corporation © 2011